The aim is to determine the radon concentration in soil
gas at a specified depth within the soil. To do this, the gas must be removed from the soil and delivered to a RAD7 without dilution by outside air. A
probe, with a hollow tube and sampling holes near the tip
may be inserted into the soil, and gas drawn up the tube,
and into the RAD7.
Questions that arise are a) how can we be sure that the gas was not being diluted by air, b) how much gas did we remove, and is that an issue, and c) from what depth was the sample being drawn? |
|
|
a. The gas may be diluted by i) air flowing down the outside of the probe tube to the sampling point, ii) leaks in the tubing from the sampling point to the RAD7, and iii) taking too much gas from a sampling point too close to the surface. Care should be taken to minimize these sources of error.
|
b. The volume of gas removed depends on the technique used to extract
it and the porosity of the soil. In practice, with average soil
porosity, the RAD7 internal pump is capable of removing soil gas at the
rate of about 0.3 - 0.5 L/min. Using the GRAB protocol, in which
the RAD7 pumps for five minutes only, the volume removed will be about
2L. In continuous monitoring mode, the RAD7 pumps for five minutes
at the start of every cycle, and then for one minute in every five. So
it removes 2L at the beginning of each cycle, and then an average of
less than 0.1L per minute after that. A flow meter on the RAD7
outlet would show how fast the RAD7 is pumping gas out of the ground.
The vacuum gauge, mounted on the probe, will indicate how hard the RAD7
pump has to work to obtain its sample of soil gas.
|
c. The actual sampling point position will depend on the probe depth,
the volume of air removed and also, perhaps, the technique used and the
probe design. After the probe is inserted, the ground should be
tamped down around the probe, to prevent air or gas from moving vertically
along the outside of the shaft. This not only prevents fresh air
from descending from the surface down to the sampling point, but also
helps to locate the gas sampling point at the position of the probe point. Assuming
that the gas is drawn equally from all directions, and assuming that
half the soil space is occupied by gas, the soil volume sampled, for
a gas sample of 2L, will be about 4L. This occupies a sphere of
radius 8cm, or 3 inches. Taking twice as much gas will increase
the sphere radius to10cm, or 4 inches.
|
CHOICE OF PROBE |
The RAD7 may be shipped with any of a number of commercially available
soil gas probes. One system, the AMS Original Gas Vapor Probe
( www.ams-samplers.com ) , comes with a 15 inch (38 cm) probe and liner
rod, a 3 ft (91 cm) extension with liner rod, a cross handle and adapter,
and a barbed
hose
adapter. This is intended for use in fairly light soils down to
a depth of no more than 4ft (1.2m). To reach this depth, it may
be necessary to use a slide hammer (available as an optional extra). With
this or any probe, care should be exercised during insertion, to avoid
the probe being damaged or bent. There is no warranty covering
damage to the probe.
With this 'Original' system, great care must be taken to ensure that the screw joints in the assembly do not leak. Teflon tape is essential, to wrap around the screw threads before assembly.
The probe tip is of wider diameter than the rest of the tubing, to keep the sampling holes unplugged during insertion. This means that extra care and effort must be exerted to tamp down the soil around the shaft.
A better system would be the AMS Retract-A-Tip, that has a tip connected by Teflon tubing to the surface. Many of the concerns associated with the Original Gas Vapor Probe, above, are eliminated, but the cost for the most basic kit is nearly ten times that of the 'Original'.
A water shup-off valve and vacuum gauge is supplied with the probe.
This is mounted on the probe head and prevents water from being sucked
up into the RAD7. The vacuum gauge will indicate a vacuum if rising water
shuts off the valve or if the soil is insufficiently porous to allow
a soil gas sample to be taken.
MEASUREMENT RESULTS
Some measurements were made with an 'Original Gas Vapor Probe' system
at a depth of 15 inches just outside the DURRIDGE offices. Initial
measurements were made in GRAB mode, to see if the values dropped from
the first reading, due to infiltration of fresh air through the ground
to the sampling point. Subsequently, the RAD7 was put into 1-day
protocol, with half-hour cycles. After an hour or so, it started
snowing, eventually covering the ground around the probe with a layer
of snow, that prevented the emanation of radon from the ground into
the air. The growth in radon concentration at 15 inches depth,
due to the snow on the surface, was monitored for a few hours.
Because there was no drop in concentration from the first grab sample
to subsequent samples, the results indicated that, even with a sampling
point as shallow as 15 inches from the surface, any dilution of the
radon concentration in the sample due to infiltration of outside air
was
insignificant, and that measuring the soil gas concentration with a
RAD7 in 1-day protocol is a valid technique.
|
 |